Customization: | Available |
---|---|
Output Signal Type: | Digital Type |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Gas Measured
|
Nitrous Oxide (N2O)
|
Cartridge Part Number
|
MIDAS-I-N2O
|
Sensor Technology
|
Nondispersive Infrared (NDIR) sensor
|
Measuring Range
|
N2O 0 - 1000ppm
|
Minimum Alarm 1 Set Point
|
125ppm
|
Lower Detection Limit
|
90ppm
|
Linearity
|
<± 10% of full scale
|
Repeatability
|
< ± 40ppm @20°C (68°F) ambient
|
Resolution
|
10ppm
|
Response Time t62.5
|
≤ 30 seconds
|
Operating Temperature
|
0°C to +40°C (32°F to 104°F)
|
Operating Humidity
|
0 to 95% RH non-condensing
|
Operating Pressure
|
90 - 110kPa
|
Effect of Position
|
No effect in typical application
|