Accuracy Specifications | ||||
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits]). | ||||
AC and DC Voltage | ||||
Function | Range | Resolution | Accuracy | |
15B MAX | 17B MAX | |||
AC Volts (40 Hz to 500 Hz)[1] K | 6.000 V 60.00 V 600.0 V 1000 V | 0.001 V 0.01 V 0.1 V 1 V | 1.0 % + 3 | 1.0 % + 3 |
AC Millivolts u | 600.0 mV | 0.1 mV | 3.0 % + 3 | 3.0 % + 3 |
DC Millivolts M | 600.0 mV | 0.1 mV | 1.0 % + 10 | 1.0 % + 10 |
DC Volts V | 6.000 V 60.00 V 600.0 V 1000 V | 0.001 V 0.01 V 0.1 V 1 V | 0.5 % + 3 | 0.5 % + 3 |
[1] All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified. | ||||
AC and DC Current | ||||
Burden voltage (typical value) AC/DC Current µA: 100 µV / µA AC/DC Current mA: 2 mV/mA AC/DC Current A: 0.03 V/A | ||||
Function | Range | Resolution | Accuracy | |
15B MAX | 17B MAX | |||
AC Current μA (40 Hz to 400 Hz) g | 400.0 μA 4000 μA | 0.1 μA 1 μA | 1.5 % + 3 | 1.5 % + 3 |
AC Current mA (40 Hz to 400 Hz) l | 40.00 mA 400.0 mA | 0.01 mA 0.1 mA | 1.5 % + 3 | 1.5 % + 3 |
AC Current A[1] (40 Hz to 400 Hz) a | 4.000 A 10.00 A | 0.001 A 0.01 A | 1.5 % + 3 | 1.5 % + 3 |
Diode Test, Temperature, Resistance, Capacitance, Frequency, and Duty Cycle | |||||
Function | Range | Resolution | Accuracy | ||
15B MAX | 17B MAX | ||||
Diode Test[1] | 2.000 V | 0.001 V | 10 % | 10 % | |
Temperature[6] T | 50.0 °C to 400.0 °C 0 °C to 50.0 °C -55.0 °C to 0 °C | 0.1 °C | NA | 2 % + 1 °C 2 °C 9 % +2 °C | |
Resistance (Ohms)[2] Ω | 400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ | 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ | 0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5 % + 3 | 0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5 % + 3 | |
Capacitance[3] E | 40.00 nF 400.0 nF 4.000μF 40.00 μF 400.0 μF 2000 μF | 0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF | 2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 | 2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 | |
Function | Range | Resolution | Accuracy | ||
15B MAX | 17B MAX | ||||
Frequency[4] (10 Hz to 100 kHz) | 50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz | 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz | NA | 0.1 % + 3 | |
Duty Cycle[4] | 1 % to 99 % | 0.1 % | NA | 1 % typical[5] | |
[1]Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA. [2]Typical open circuit test voltage: 0.54 V. Maximum short circuit current: 1.8 mA. [3]Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 40 nF range). [4]All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Input below 1 % of range are not specified. [5]Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %. [6]Use type K thermocouple. |
Continuity Threshold | ||||
Function | Threshold | |||
Continuity Threshold | 70 Ω | |||
Function | Overload Protection | Input Impedance (Nominal) | Common Mode Rejection Ratio | Normal Mode Rejection Ratio |
AC Volts | 1000 V[1] | >10 MΩ, <100 pF | >60 dB at 50 Hz or 60 Hz | - |
AC Millivolts | 1000 V[1] | >1 MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
DC Volts | 1000 V[1] | >10 MΩ, <100 pF | >100 dB at 50 Hz or 60 Hz | >60 dB at 50 Hz or 60 Hz |
DC Millivolts | 1000 V[1] | >1 MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |